NI 4071 7½-Digit FlexDMM Calibration Procedure
47. Call the niDMM Cal Adjust Gain VI with the following parameters:
48. Generate 10 k
on the calibrator with external sense turned on and 2-wire compensation
turned off.
49. Call the niDMM Cal Adjust Gain VI with the following parameters:
50. Generate 0
on the calibrator with external sense turned on and 2-wire compensation
turned off.
51. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
100e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
The value that is
displayed on the calibrator for 0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
10e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
The value that is
displayed on the calibrator for 10 k