C.A 6165 

 

 

 
 
Appliance multitester

 

 

GB - 

Quick start guide

 

Содержание C.A 6165

Страница 1: ...C A 6165 Appliance multitester GB Quick start guide ...

Страница 2: ...ures 7 3 3 Symbols and messages 8 4 Single tests 11 4 1 Single test measurements 11 4 1 1 Visual inspections 11 4 1 2 Continuity 12 4 1 3 HV AC 15 4 1 4 HV DC 16 4 1 5 HV AC programmable 18 4 1 6 HV DC programmable 19 4 1 7 Insulation resistance Riso Riso S 21 4 1 8 Sub leakage Isub Isub S 23 4 1 9 Differential Leakage 25 4 1 10 Ipe Leakage 27 4 1 11 Touch Leakage 29 4 1 12 Power 30 4 1 13 Leak s ...

Страница 3: ...es Use only standard or optional test accessories supplied by your distributor Only test adapters provided or approved by Chauvin Arnoux should be connected to TC1 test and communication connectors Use only earthed mains outlets to supply the instrument In case a fuse has blown refer to chapter 5 1 Fuses in this Instruction manual to replace it Instrument servicing and calibration is allowed to be...

Страница 4: ...ooling is required before proceeding with tests Maximum intermittent duty cycle for measurements with load currents higher than 10 A is 50 1 1 2 3 Insulation resistance Do not touch the test object during the measurement or before it is fully discharged Risk of electric shock 1 1 3 Markings on the instrument Read the Instruction manual with special care to safety operation The symbol requires an a...

Страница 5: ...for external test adapters 6 Mains test socket 7 P S probe connector 8 Keypad 9 HV output connectors 10 HV output warning lamp 11 Continuity connectors 12 Insulation Subleakage connectors 13 Discharging time connectors 14 Colour TFT display with touch screen 15 Control outputs 16 Control inputs 17 Multipurpose RS232 1 port 18 Multipurpose RS232 2 port 19 Ethernet connector C A 6165 ...

Страница 6: ...6 20 USB connector 21 MicroSD card slot ...

Страница 7: ...d view of options HV Test key is used to start and stop HV tests 3 2 General meaning of touch gestures Tap briefly touch surface with fingertip is used to select appropriate option confirm selected option start and stop measurements Swipe press move lift up down is used to scroll content in same level navigate between views in same level long Long press touch surface with fingertip for at least 1 ...

Страница 8: ...f the too high current is only of short duration caused by a short inrush current the test can be performed otherwise not Select YES to proceed with or NO to cancel measurement Resistance L N 30 Ω In pre test a low input resistance of the device under test was measured This can result in a high current after applying power to the device If the high current is only of short duration caused by a sho...

Страница 9: ...rted Press OK to continue The load current continuously exceeded 10 A for more than 4 min moving average in Power and Leakage tests Measurement was stopped for safety Press OK to continue The load current exceeded the highest upper limit of 16 A for the Power and Leakage tests Measurement was aborted Press OK to continue Warning for restart of the instrument to set new Ethernet settings This messa...

Страница 10: ...he instrument output Withstanding test voltage Insulation test voltage or mains voltage Warning A very high and dangerous voltage is will be present on the instrument output Withstanding test voltage Test passed Test failed Conditions on the input terminals allow starting the measurement consider other displayed warnings and messages Conditions on the input terminals do not allow starting the meas...

Страница 11: ...tion menu Test circuit Figure 4 2 Visual inspection test circuit Visual inspection procedure Select the appropriate Visual inspection Start the inspection Perform the visual inspection of the appliance equipment Apply appropriate ticker s to items of inspection End inspection Save results optional ...

Страница 12: ... A Test parameters Output connections Output 4 wire P PE Test current I out 0 2 A 4 A 10 A 25 A Duration Duration Off 2 s 180 s ΔU test Enables ΔU test On Off Wire section Wire section for ΔU test 0 5 mm2 6mm2 Test limits H Limit R H limit Off 0 01 Ω 9 Ω Custom L Limit R L limit Off 0 01 Ω 9 Ω Custom H Limit ΔU H limit 1 0 V 5 0 V Applicable only at test current 10 A ...

Страница 13: ... parameters limits Connect test leads to C1 P1 P2 and C2 terminals on the instrument 4 wire or connect test lead to P S terminal 2 wire measurement P S PE Compensate test leads resistance optional Connect test leads to device under test Start measurement Measurement can be stopped manually or by timer Save results optional Figure 4 7 Examples of Continuity measurement results ...

Страница 14: ...n for compensating the resistance of test leads Figure 4 8 Shorted test leads Compensation of test leads resistance procedure Select the Continuity function Parameter Output must be set to P S PE Connect test leads to the instrument and short the test leads together see Figure 4 8 Touch the key to compensate leads resistance Symbol is displayed if the compensation was carried out successfully Figu...

Страница 15: ...st results sub results I test current U measured a c test voltage Ir resistive portion of test current Ic capacitive portion of test current Test parameters AC test voltage U test 100 V 5000 V in steps of 10 V Duration t end Off 1 s 120 s Test limits High limit I H limit 0 5 mA 100 mA Low limit I L limit Off 0 5 mA 100 mA Test circuit Figure 4 11 HV AC measurement ...

Страница 16: ...sults Note First HV measurement after power on the instrument if password protection is enabled or first HV measurement after enabling or changing password require entering password for enabling HV test 4 1 4 HV DC IMPORTANT SAFETY NOTE Refer to chapter 1 1 Warnings and notes for more information regarding safe use of the instrument Figure 4 13 HV DC test menu Test results sub results U measured t...

Страница 17: ...test leads to HV and HV terminals on the instrument Connect HV test leads to device under test Start measurement Measurement can be stopped manually or by timer Save results optional Figure 4 15 Examples of HV DC measurement results Note First HV measurement after power on the instrument if password protection is enabled or first HV measurement after enabling or changing password require entering ...

Страница 18: ...st Figure 4 17 HV AC programmable test menu Test results sub results I test current U measured test voltage Ir resistive portion of test current Ic capacitive portion of test current Test parameters Starting AC test voltage U start 100 V 5000 V in steps of 10 V AC test voltage U test 100 V 5000 V in steps of 10 V Duration of starting voltage t start 1 s 120 s Duration of ramp t ramp 2 s 60 s Durat...

Страница 19: ... results optional Figure 4 19 Examples of HV AC programmable test results Note First HV measurement after power on the instrument if password protection is enabled or first HV measurement after enabling or changing password require entering password for enabling HV test 4 1 6 HV DC programmable IMPORTANT SAFETY NOTE Refer to chapter 1 1 Warnings and notes for more information regarding safe use of...

Страница 20: ...parameters Starting DC test voltage U start 500 V 6000 V in steps of 50 V DC test voltage U test 500 V 6000 V in steps of 50 V Duration of starting voltage t start 1 s 120 s Duration of ramp t ramp 2 s 60 s Duration of test voltage t end Off 1 s 120 s Test limits High limit I H limit 0 05 mA 10 0 mA Low limit I L limit Off 0 05 mA 10 0 mA Test circuit Figure 4 21 HV DC programmable test ...

Страница 21: ...ing or changing password require entering password for enabling HV test 4 1 7 Insulation resistance Riso Riso S Figure 4 23 Insulation resistance test menus Test results sub results Riso Insulation resistance Riso S Insulation resistance S Um Test voltage Test parameters Nominal test voltage Uiso 50 V 100 V 250 V 500 V 1000 V Duration Duration Off 2 s 180 s Type of test Type Riso Riso S Riso Riso ...

Страница 22: ...22 Test circuits Figure 4 24 Measurement of insulation resistance ISO ISO Figure 4 25 Measurement of insulation resistance Socket LN PE Figure 4 26 Measurement of Riso Riso S socket ...

Страница 23: ...sults optional Figure 4 27 Examples of Insulation resistance measurement results Note When P S probe is connected during the Riso measurement then the current through it is also considered 4 1 8 Sub leakage Isub Isub S Figure 4 28 Sub Leakage test menus Test results sub results Isub Sub leakage current Isub S Sub leakage current S Test parameters Type of test Type Isub Isub S Isub Isub S Output vo...

Страница 24: ...re 4 29 Measurement of Sub leakage SUB1 SUB2 Figure 4 30 Measurement of Sub leakage socket LN PE Figure 4 31 Measurement of Sub leakage Sub leakage S socket Sub leakage measurement procedure Select the Sub leakage function Set test parameters limits Connect test leads to SUB1 SUB2 terminals on the instrument then connect test leads to device under test or ...

Страница 25: ...t through it is also considered 4 1 9 Differential Leakage Figure 4 33 Differential Leakage test menu Test results sub results Idiff Differential Leakage current P Power Test parameters Duration Duration Off 2 s 180 s Change status Change YES NO YES The instrument measures leakage current in two sequential steps with delay in between The phase voltage is firstly applied to the right live output of...

Страница 26: ...26 Test limits H Limit Idiff H limit Off 0 25 mA 15 0 mA Custom L Limit Idiff L limit Off 0 25 mA 15 0 mA Custom Output connections Socket L N PE P S ...

Страница 27: ...ential Leakage function Set test parameters limits Connect device under test to mains test socket and optionally to P S terminal Start measurement Measurement can be stopped manually or by timer Save results optional Figure 4 35 Examples of Differential Leakage measurement results 4 1 10 Ipe Leakage Figure 4 36 Ipe Leakage test menu ...

Страница 28: ...s test socket NO The phase voltage is applied only to the right live output of the mains test socket Delay time Delay 0 2 s 5 s Output connections Socket L N PE Test limits H Limit Ipe H limit Off 0 25 mA 15 0 mA Custom L Limit Ipe L limit Off 0 25 mA 15 0 mA Custom Test circuit Figure 4 37 Measurement of Ipe Leakage current Ipe Leakage measurement procedure Select the Ipe Leakage function Set tes...

Страница 29: ...ent measures leakage current in two sequential steps with delay in between The phase voltage is firstly applied to the right live output of the mains test socket and secondly to the left live output of the mains test socket NO The phase voltage is applied only to the right live output of the mains test socket Delay time Delay 0 2 s 5 s Output connections Socket L N PE P S Test limits H Limit Itou ...

Страница 30: ...test parameters limits Connect device under test to mains test socket Connect test lead to P S terminal on the instrument and on device under test Start measurement Measurement can be stopped manually or by timer Save results optional Figure 4 41 Examples of Touch Leakage measurement results 4 1 12 Power Figure 4 42 Power measurement menu ...

Страница 31: ...f 2 s 180 s Output connections Socket L N Test limits H Limit P H limit Off 10 W 3 50 kW Custom L Limit P L limit Off 10 W 3 50 kW Custom Test circuit Figure 4 43 Measurement of Power Power measurement procedure Select the Power function Set test parameters limits Connect device under test to mains test socket Start measurement Measurement can be stopped manually or by timer Save results optional ...

Страница 32: ...akage current in two sequential steps with delay in between The phase voltage is firstly applied to the right live output of the mains test socket and secondly to the left live output of the mains test socket NO The phase voltage is applied only to the right live output of the mains test socket Delay time Delay 0 2 s 5 s Output connections Socket L N Socket L N PE P Test limits H Limit P H limit O...

Страница 33: ...cedure Select the Leak s Power function Set test parameters limits Connect device under test to mains test socket and optionally to P S terminal Start measurement Measurement can be stopped manually or by timer Save results optional Figure 4 47 Examples of Leak s Power measurement results ...

Страница 34: ...rging time function is as following Phase The device under test is connected to supply voltage via an external socket The instrument monitors the voltage on supply or internal connections and internally stores the peak voltage value Phase The device under test is disconnected from the supply and the voltage at the test terminals starts to fall Once the rms voltage falls for 10 V the instrument sta...

Страница 35: ...g Time test Output External Discharging Time test procedure Output External Select the Discharging Time function Set test parameters limits Connect test leads to the DISCHARGING TIME terminals on the instrument and on the device under test Connect device under test to the mains supply and Switch it ON Start measurement Measurement is stopped manually by disconnecting device under test mains supply...

Страница 36: ...ects the DEVICE UNDER TEST from the supply and the voltage at the supply connections starts to fall Disconnection moment is always at peak voltage Phase After the voltage drops below the limit value the timer is stopped Test circuit Output Socket Figure 4 52 Discharging Time test Output Socket Discharging Time test procedure Output Socket Select the Discharging Time function Set test parameters li...

Страница 37: ...parameters and limits are the same as set in the Power single test see chapter 4 1 12 Power Test circuit Figure 4 55 Functional inspection Functional inspection procedure Select the appropriate Functional inspection Start the inspection Perform the functional inspection of the appliance equipment Perform the Power measurement test through the mains test socket optional Apply appropriate ticker s t...

Страница 38: ...38 Figure 4 56 Examples of Functional Inspection results ...

Страница 39: ... with the same types defined in this document 5 2 Warranty Except as otherwise stated our warranty is valid for 24 months starting from the date on which the equipment was sold Extract from our General Conditions of Sale provided on request The warranty does not apply in the following cases Inappropriate use of the equipment or use with incompatible equipment Modifications made to the equipment wi...

Страница 40: ...695739A02 Ed 1 03 2018 ...