Programming
Chapter 5
AT-MIO-64F-5 User Manual
5-6
© National Instruments Corporation
In posttrigger sequences, the sample counter starts counting after receipt of the first trigger, while
in the pretrigger acquisition mode, the sample counter does not start counting until a second
trigger condition occurs. The data acquisition operation is initiated by writing to the DAQ Start
Register or by a falling edge on the EXTTRIG* signal. Programming multiple A/D conversions
on a single channel requires the following programming steps for posttrigger and pretrigger
modes, as well as internal and external timing. The instructions in the blocks of the following
flow chart are enumerated in the
Data Acquisition Programming Functions
section later in this
chapter.
START
Program a single
analog input channel,
gain, mode, and range
Program the
sample-interval
counter
Program the
sample counter
Clear the A/D
circuitry
Enable a single-channel
data acquisition
operation
END
Apply a trigger
Service the data
acquisition operation
Figure 5-3. Single-Channel Data Acquisition Programming
Summary of Contents for AT-MIO-64F-5
Page 13: ......
Page 16: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......