CellScale
®
System • 3750CS-5 User Guide Page 83
MEASUREMENT SYSTEMS INTERNATIONAL
Software V2-21 or greater, 2nd Generation Modem
To ENABLE / DISABLE AZM
1
...
1
AZM
2
ENABLED
3
DISABLED
or
ENABLED
4
AZMRANGE
5
1_
5
10_
6
AZM
7
or
1) Use the UP or DOWN scroll key to locate the menu
choice “AZM”.
2) Push ENTER to access the AZM submenu.
3) Change AZM to “DISABLED” or “ENABLED” as
desired using the UP or DOWN scroll key.
4) Push ENTER.
5) If you selected “ENABLED” in the previous step
the “AZMRANGE” screen appears. Use the numeric
keys to input number of scale divisions the system
can zero out automatically. Use “0” for standard AZM
operation. Using “0” sets the AZM window at ±0.5d.
The range of AZM is limited to 255d.
In this example, the scale is a 1000 kg x 0.1 kg. Therefore to achieve an
example AZM range of 1 kg, we’ll enter “10” (10 x 0.1 kg = 1 kg).
6) After keying in the desired AZM range, push
ENTER.
7) Push EXIT to return to normal scale operation, or
push UP or DOWN to choose another Calibrate Setup
function.
MOTION BAND
The motion band determines the range of weight variation that the indicator considers to be stable weight. Since
the CellScale will not zero or tare a value in motion, the motion band directly influences the accuracy of the scale
system. The motion band in Legal-For-Trade systems is fixed at ≤ 0.6d and cannot be altered. In systems used
for industrial applications, the motion band can be adjusted larger. Adjusting the motion band larger can help in
medium accuracy applications where there is a lot of vibration or noise and it is difficult to hold the weight steady
enough to tare, zero, or total. For example, setting the motion band to 5 d would allow a tare (or zero or total) to
be as much as 5d off of its ideal value but it will make the indicator easier to use as you don’t have to wait until
the weight is completely stable.
The Motion Band ± range is limited to a maximum of 255 `d`. If you exceed this range, the 3750 will return the error message “TOO BIG”. Setting the Motion Band
to 0 is equal to a range of ±0.5d.
To Adjust the Motion Band
Before starting the following procedure, you must be in the Top Menu Level of Calibration. See “Enable Calibration”.
1
...
1
MOTNBAND
2
1
MOTNBAND
3
2_
4
MOTNBAND
5
or
1) Use the UP or DOWN scroll key to scroll to the
“MOTNBAND” message.
2) Press ENTER to access the Motion Band Submenu.
The current motion band value is displayed on the
numeric digits.
3) The “MOTNBAND” screen will appear. Use the
numeric keys to input the weight range the system
uses to detect Motion. Use “1” for standard operation.
This sets the Motion Band to ±1d.
In this example, we’ll enter ±2d as the Motion Band.
4) After entering the desired motion band, push
ENTER.
5) Push EXIT to return to normal scale operation, or
push UP or DOWN to choose another Calibrate Setup
function.