CALIBRATE SETUP MENU
LEGAL STANDARD
The STANDARD function allows the Model 3750CS to be configured as required by various regulatory agencies.
The “INDUSTRY” selection is the default. Choose “NIST” for US and Canadian Legal-for-Trade units. Choose
“OIML” for European L-F-T applications. “METRIC” is the same as “INDUSTRY” except only kg and g units
are enabled.
Before starting the following procedure, you must be in the Top Menu Level of Calibration. See “Enable Calibration”. Contact MSI for details on the differences
between the various configurations.
To Select the Legal Standard
1
...
1
STANDARD
2
INDUSTRY
3
NIST
3
OIML
3
METRIC
4
STANDARD
5
or
1) Use the UP or DOWN scroll key to scroll to the
“STANDARD” message.
2) Press ENTER. The current Standard is displayed.
3) Use the UP or DOWN key to scroll through the
Standards options. For this example we’ll pick the
“METRIC” standard. This will limit the units choices
to kg and g (and Metric Tons if the capacity is high
enough).
4) Push ENTER to set your selection.
5) You are now back in the “CAL SETUP” menu. Use
the UP/DOWN keys to scroll to the next calibration
operation or push EXIT to return to normal scale
operation.