Configuring the RF Modem
Assumes F1 is configured as “MODEM”, see “Setup Function Keys”.
1) Push the SETUP key.
2) Push the MODEM key (usually F1). The display
enters the Setup MODEM mode.
3) Push ENTER. The first menu item is RF NTWK.
The numeric digits indicate the current network.
4) Change the Network number by pushing the numeric
keys. As soon as you start entering numbers the
screen reads “NEW N/W”. Available network num-
bers range from 0-31.
In this example we’ll use network 12. Use the appropriate numbers for your
CellScale master. It is not necessary to enter a ‘0’ for networks below 10.
Push the 1 key followed by the 2 key.
5) Push ENTER to set your Network choice into
memory. The 3750CS tries to contact the new net-
work while indicating “FINDING”. If the master
unit is on it will connect. If not, the display will flash
briefly “NO NETWK”.
The “NO NETWK” or “NO TARGET” messages can be safely ignored at this
time. It indicates only that the chosen network is not on or in range.
6) Push the UP key to the “C/S ID” menu.
7) Use the numeric keys to change the C/S ID number
to match the SRC address programmed in the
CellScale master. Available C/S ID numbers range
from 0-254.
In this example we’ll use source 75. Use the appropriate numbers for your
CellScale master. It is not necessary to enter leading zeros.
8) Push the ENTER key to set the C/S ID.
9) Push the UP key to the “MY ID” menu. This is the
Node ID of this 3750CS remote.
10) This step is optional. My ID is important only if mul-
tiple slaves are present on the set Network. Use the
Numeric keys to input a unique ID for this 3750CS,
then push ENTER.
11) Push the UP key to the “ENERGY” menu. The
numeric display indicates the current energy level, 2
equals High Power, 1 equals Med Power, 0 indicates
Low power.
12) Press the 0, 1 or 2 key as needed.
MSI recommends using high power (2) in most
installations.
13) This completes the MODEM setup. Push the EXIT
key to return to normal scale operation.
14) If the 3750CS is reporting “NO TARGET” it may
be necessary to send the “Talk to me” message. Do
this by pushing the MODEM key (usually F1).
Enter ID for this 3750CS
NEW ID
2
ENERGY