Page 12 MSI-3750CS-5 RF Weighmeter • User Guide
3750CS
C
ELL
S
CALE RF DIGITAL WEIGHT INDICATOR
®
The 3750CS serves as a terminal for CellScales. Its response to key presses may be delayed due to transmission
times. The 3750CS will on occasion, display a “WAIT” message while transacting setups and scale controls.
!
RF NETWORK SETUP
The 3750CS is a RF Modem connected device. The RF Modem requires setup to connect to one or more
CellScales.
0=LOW
1=MED
2=HIGH
3750CS RF ENERGY
Enter
0-255
Enter
0-63
MODEM SETUP MENU
RF NTWK
ENERGY
C/S ID
NETWORK 0-63
Enter
0-254
3750CS
NODE ID
MASTER
CellScale ID
MY ID
EXIT
SETUP
!
MODEM
F1
Enter
0-1800s
TIMEOUT
Any F key
Programmed
to “Modem”
RF NETWORK
Sets the primary network number. Set this numbers to match the Network number of the CellScale master the
3750CS is associated with most often. Ranges from 0 to 63.
C/S ID
Sets the primary master CellScale ID number. This number must match the Source Address number of the
primary master CellScale. Ranges from 0 to 254.
MY ID
Each remote unit in a given Network must have a unique ID number. Enter up to three digits to designate the
3750CS ID number. Ranges from 0 to 254.
TIMEOUT
Sets the time in seconds (0-1800) that the 3750CS will wait for data from the CellScale. The display will blank
out if no data is received within this time period. Defaults to 10 seconds. Entering 0 turns off the Time-out
timer. MSI recommends at least a 5 second timeout period to handle small lapses in transmissions. The system
defaults to 10 seconds.
ENERGY
Sets the RF power level between 10mW (Low), 63mW (Med) and 250mW (High). In most applications power
should be set to High. Reduced energy can reduce the size of the coverage/interference zone which may be
desirable for multiple-network applications. Always check for adequate range coverage when using the Low
or Med setting.