3.5.5
ELEVATE/REACH SEAT FUNCTION - THRU SWITCH
ELEVATE/REACH INSTRUCTIONS:
(
Note:
Elevate(Reach) operates in
DIRECT
mode thru the 8-way Switch).
1. Press backward on the ELEVATE switch (
1
) to elevate the seating
system upward
.
2. Press forward on the ELEVATE switch (
2
) to lower the seating
system downward.
1
2
3.5.6
ANTERIOR TILT FUNCTION - THRU SWITCH
ANTERIOR TILT INSTRUCTIONS:
(
Note:
Anterior-Tilt operates in
DIRECT
mode thru the 8-way Switch).
1. Press backward on the ‘A TILT’ switch (
1
) to raise the anterior tilt
function (seating system) forward to the full Reach position
.
2. Press forward on the ‘A TILT’ switch (
2
) to lower the anterior tilt
(seating system) downward.
1
2
The Rovi-Upfront MPPS uses a combination of mechanical switches and a programmable Tipsy Angle
Sensor to set all the required safety limits and lockouts for the wheelchair. This includes programmed
settings for the Home position, the Maximum Tilt (or Tilt/Recline) Limit, the Drive Lockout Limit and the
Reduced Drive Speed Limit. Refer to
Section 4.3.6 Tipsy Angle Sensor
for more detailed information
on the recommended safety limit settings.
When operating your Rovi-Upfront MPPS thru an 8-way switch, the LED indicator lights on the switch
control box provide a visible indicator to notify if/when the wheelchair has reached the established safety
limits for
Reduced Drive Speed
(
RDS
) and/or
Drive Lockout
(
DLO
):
3.5.7
THRU SWITCH, SAFETY INDICATORS
(
FOR
REDUCED DRIVE SPEED
&
DLO):
REDUCED DRIVE SPEED (RDS)
:
The RDS Limit places the wheelchair into a reduced drive speed
(prior to DLO) when the seating system is tilted and/or reclined beyond
a preset angle, or raised beyond the ‘home” position for the Anterior-
Tilt and/or Elevate/Reach function. When the Rovi Upfront MPPS
is in reduced drive speed, the LED indicator lights are “
Flashing RED
“.
I.
S
WITCH
I
NDICATORS FOR
RDS
AND
DLO
8-WAY SWITCH CONTROL
DRIVE LOCKOUT
(
DLO
):
The DLO Limit prevents the wheelchair from driving when the seating
system is tilted
(anterior or posterior tilted)
and/or reclined beyond a
preset (safe) operating angle. When the Rovi Upfront MPPS is in drive
lockout, the LED indicator lights are “
Solid RED
”.
LED indicator lights
*NOTE:
In order to return the wheelchair to normal drive speed, the affected power positioning function
must be returned to its established ‘Home’ position. (When the ROVI Upfront MPPS is in the ‘Home’
position, the LED indicator lights remain off).
- 38 -
3.0 PROGRAMMING
3.0 PROGRAMMING
AND
AND
OPERATING
OPERATING
YOUR
YOUR
ROVI-UPFRONT MPPS
ROVI-UPFRONT MPPS
Summary of Contents for ULTRA-LOW MAXX UPFRONT MPPS