- 36 -
IMPORTANT!
The RNet Remote Joystick MUST be powered on to enable the 8-Way Switch Control.
When a separate (8-way) switch control is installed on the Rovi-Upfront power positioning system, the seating
functions may be operated independent from the joystick. (The RNet remote joystick does not have to be
placed into ‘Power Seating Mode’ in order to operate the power positioning functions; however the joystick
power must be turned on).
The available power seating functions are labeled on the 8-way switch control. To operate a selected power
positioning function, press and hold the rocker switch (button) FORWARD or BACKWARD to activate the
actuator motor (seat function). The actuator motor will stop when the rocker switch/button is released. The
switch control can be independently programmed
*
to meet the preferences of the end user. The direction of
the rocker switch and the corresponding seat function (e.g.; tilt up/tilt down, elevate up/elevate down, etc...) will
be dependent on whether the power seating function (rocker switch) is programmed to operate in direct-
mode or toggle-mode:
i)
Direct Mode:
two switch/button directions are required to operate a designated power seating function,
one for each actuator motor direction.
ii)
Toggle Mode:
only one switch/button direction is required for each power seating function. In this mode,
pressing the switch once will activate the seat function (actuator motor); releasing the switch (pausing 2 sec.),
then pressing the same switch again will cause the actuator motor to travel in the reverse direction.
3.5 Operating the ROVI-UPFRONT MPPS - Thru a Separate Switch Control
= TILT
(DIRECT mode):
•
Tilt Up-
Press Switch Forward;
•
Tilt Down-
Press Switch Backward
= RECLINE
(TOGGLE mode):
•
Recline Back-
Press Switch Forward;
•
Recline Up-
Release, Pause, then Press Switch Forward again
= POWER LEGS
(TOGGLE mode):
•
Legrest Extend-
Press Switch Backward;
•
Legrest Retract-
Release, Pause, then Press Switch Backward
= ELEVATE
(DIRECT mode):
•
Elevate Up-
Press Switch Forward;
•
Elevate Down-
Press Switch Backward
= ANTERIOR TILT
(
REACH
) (DIRECT mode):
•
Anteriror Tilt (Reach) Up-
Press Switch Forward;
•
Anteriror Tilt (Reach) Down-
Press Switch Backward
8-WAY SWITCH CONTROL
The rocker switches on the 8-way Switch Control may be programmed to operate in either Direct Mode and/or
Toggle Mode depending on the seating function to which they are assigned. The following image illustrates
the standard (factory) switch configuration for the Rovi Upfront MPPS, and a basic functional overview of the
rocker switches (push buttons)
1
2
3
4
8-WAY SWITCH CONTROL
1
4
3
2
3.5.1
SEAT CONTROLS THRU SWITCH
5
5
rocker
switch
LED indicator lights
*Note
: The 8-way switch control on your Rovi Upfront MPPS has been pre-programmed; Programming
changes must only be performed by a Qualified Service Technician. If you require any changes to your existing
switch configuration, please contact your local Service Provider for assistance.
IMPORTANT!
Be certain to read through
Section 3.3,
Section 3.3,
Understanding the Power Positioning
Understanding the Power Positioning
Functions
Functions, BEFORE operating your Rovi Upfront MPPS for the first time.
3.0 PROGRAMMING
3.0 PROGRAMMING
AND
AND
OPERATING
OPERATING
YOUR
YOUR
ROVI-UPFRONT MPPS
ROVI-UPFRONT MPPS
Summary of Contents for ULTRA-LOW MAXX UPFRONT MPPS