M
AVTM246004 BITE 2/2P Rev 4 July 2011
44
Retesting Cells and
Straps
You cannot review a test after it
has been completed. (Refer to
"Step Seven: Measuring the Cell
and the Strap," #'s 6,7, and 8,
for a description of completing a
test.)
If needed, you can retest any of the cells or straps in the
current test.
1.
Navigate through the test screens until information about
the cell or strap you want to retest is displayed.
2.
Pull the trigger.
The
receiver
enters into test mode.
001
XX.XX
XX.XX
T01
3.
Position the
receiver
and
potential probe
on top of the
battery cell terminals, and then press down.
4.
When the voltage and impedance values displayed on the
LCD stabilize, pull the
trigger
to store the readings.
See the sample placement
shown in Figure 4-7 on page 40.
5.
When the strap values displayed on the screen stabilize,
pull the
trigger
to store the reading.
NOTE
:
Both measurements must be taken.
6.
To return to testing, scroll forward to the next cell.
7.
To retest another cell/strap, repeat steps 1-5.
Printing an Active
Test to Review the
Data (BITE 2P)
To print the test data of the string for which you are taking
measurements, disconnect the
potential probe
and connect
the printer cable to J3. The LCD will display the following
screen.
www.
GlobalTestSupply
.com
Find Quality Products Online at: