AVTM246004 BITE 2/2P Rev 4 July 2011
32
Step Four: Setting Baseline, Warning and Fail Values
The BITE 2/2P now lets you set baseline, warning and fail
impedance values when conducting a test. Then, after
displaying the cell or strap measurements, the BITE 2/2P
receiver
screen will indicate PASS, FAIL or WARN and will
display the measured value as a percentage of the baseline
value.
After you finish wanding information, the
receiver
will display
the following menu:
Y
ENTER BASELINE?
N
(Y/N)
MENU
If you do not want to set baseline, warn and fail values, press
(for NO). The receiver screen will ask whether you want to
split the strap. Go to "Step Five: Splitting the Strap below.
If you want to enter baseline, warn and fail values, press
(for YES).
BASE
WARN
FAIL
XX.XX
XX%
XX%
1.
Use the
to scroll through the digits until the desired
digit is displayed.
2.
Press
.
3.
Repeat for each digit and the decimal point until the
desired baseline, warn and fail values are displayed.