Pressure Support
ressure Support
essure Support
ssure Support
sure Support
ure Support
re Support
e Support
Support
Support
upport
pport
port
ort
rt
t
Pressure Support (NIV) is a patient initiated
ure Support (NIV) is a patient initiated
breathing mode in which the ventilator
supports the patient with a set constant
upports the patient with a set constant
pressure.
The following parameters are set:
meters are set:
1. PS (Pressure Support level) above PEEP
bove PEEP
(cmH
2
O)
2. PEEP (cmH
2
O)
3. Oxygen concentration (%))
4. Inspiratory rise time (s)
5. Inspiratory Cycle-off (%)
cle-off (%)
6. NIV rate (b/min)
/min)
7. Backup Ti (s)
Differences from invasive Pressure
m invasive Pressure
support mode:
ort mode:
• When the Standby key is pressed a
waiting position dialog is shown. All
osition dialog is shown. All
patient related alarms are turned off during
urned off during
120 seconds. Press the Start ventilation
pad to start the ventilation.
• During NIV the ventilator automatically
NIV the ventilator automatically
adapts to the variation of leakage in order
dapts to the variation of leakage in order
to maintain the required pressure and
quired pressure and
PEEP level. If the leakage is excessive, the
kage is excessive, the
ventilator will issue a high priority alarm,
y alarm,
deliver a continuous flow and pause
pause
breath cycling. Ventilation will resume
will resume
automatically if the leakage decreases.
decreases.
Ventilation can also be started manually by
pressing the
g the Start ventilation pad in the
excessive leakage dialog.
• During Pressure support the system
ensures a minimum Back-up Rate and
Back-up Rate and
maintains the set Inspiratory pressure and
ratory pressure and
PEEP level. The Back-up Rate is activated
he Back-up Rate is activated
when the spontaneous breathing rate is
n the spontaneous breathing rate is
lower then the Back-up Rate, but the
k-up Rate, but the
ventilator does not activate a Backup
ctivate a Backup
ventilation mode as in Invasive Pressure
Pressure
Support.
• Trigger sensitivity cannot be set in NIV.
cannot be set in NIV.
Read more about NIV
ead more about NIV
ad more about NIV
d more about NIV
more about NIV
more about NIV
ore about NIV
re about NIV
e about NIV
about NIV
about NIV
bout NIV
out NIV
ut NIV
t NIV
NIV
NIV
IV
V
Intended population
page 4
4
NIV general information:
page 61
1
Ventilation modes (NIV):
page 62
ge 62
Alarm settings:
page 73
Preparation:
page 160
SVX-9014_XX
NIV - Pressure Support