1. Patient Unit: Lift the locking handle (a)
Lift the locking handle (a)
and pull out (b).
ull out (b).
2. Hinge the Expiratory cassette and press
ratory cassette and press
it firmly down into lock position. Check
Check
that the cassette cannot be moved
upwards.
pwards.
3. Patient Unit: Lift the locking handle.
he locking handle.
4. Push back the unit until you hear a
the unit until you hear a
"click".
Caution: The Expiratory cassette must be
ssette must be
pressed all the way down into its bottom
position. Make sure that the button on the
Make sure that the button on the
top of the cassette is completely ejected to
mpletely ejected to
ensure that the cassette is firmly locked.
ked.
Important:
portant:
• Note on a log sheet that a routine cleaning
has been performed. Refer to hospital
performed. Refer to hospital
guidelines.
Function test
unction test
nction test
ction test
tion test
ion test
on test
n test
test
test
est
stt
After cleaning, always perform a Pre-use
m a Pre-use
check. For more information please refer to
please refer to
page 146.