186
86
Servo
…
User´s manual
US edition
Order No: 66 00 261
7
Infant
Adult
Universal
Options
Servo Ultra Nebulizer
ervo Ultra Nebulizer
rvo Ultra Nebulizer
vo Ultra Nebulizer
o Ultra Nebulizer
Ultra Nebulizer
Ultra Nebulizer
ltra Nebulizer
tra Nebulizer
ra Nebulizer
a Nebulizer
Nebulizer
Nebulizer
ebulizer
bulizer
ulizer
lizer
izer
zer
err
WARNINGS!
GS!
• Servo Humidifier/HME must be
ME must be
disconnected during nebulization.
ation.
Otherwise the humidifier may be blocked.
be blocked.
• The heated humidifier must be switched
witched
off during nebulization. Otherwise the
Otherwise the
particle size may be affected.
ffected.
• When a Servo Ultra Nebulizer is used,
Nebulizer is used,
always consult the drug manufacturer
he drug manufacturer
regarding the appropriateness of
ultrasonic nebulization for certain
bulization for certain
medications.
• The nebulizer must not be used without
he nebulizer must not be used without
buffer liquid (sterile water). Otherwise the
(sterile water). Otherwise the
ultrasonic generator crystal may break.
ystal may break.
• Continuously check that the buffer liquid
heck that the buffer liquid
level is between MIN. and MAX. during
MIN. and MAX. during
nebulization.
tion.
• During nebulization: Continuously check
check
that moisture is generated in the
medication cup.
• During nebulization a filter must be
connected to the expiratory inlet of the
d to the expiratory inlet of the
ventilator. Always carefully monitor the
s carefully monitor the
airway pressure during nebulization.
during nebulization.
Increased airway pressure could be
pressure could be
caused by a clogged filter. The filter should
he filter should
be replaced if the expiratory resistance
increases or every 24 hours when the
very 24 hours when the
nebulizer is being used.
g used.
• When the ventilator is running on batteries
the nebulizer module is inoperative to
zer module is inoperative to
reduce the power consumption..
Cautions:
• Check that the medication cup is
undamaged and that it is firmly in place
ged and that it is firmly in place
before the nebulizer is started.
ore the nebulizer is started.
• If a nebulizer and a CO
O
2
analyzer are in use
simultaneously, the CO
2
reading may be
ding may be
affected.
Important:The Servo Ultra Nebulizer may be
rvo Ultra Nebulizer may be
interrupted shortly due to overheating. It will
d shortly due to overheating. It will
automatically start again when the buffer
cally start again when the buffer
water has cooled. During this short period of
d. During this short period of
time no alarm is activated and the timer is not
ctivated and the timer is not
interrupted.
Note: For information about the stand alone
bout the stand alone
Aeroneb Professional Nebulizer System,
zer System,
refer to separate manual.
Note: The
he
Y Sensor measurement can be
incorrect when the Aeroneb Professional
when the Aeroneb Professional
Nebulizer System is in use. Therefore, we
zer System is in use. Therefore, we
recommend that the Y Sensor is removed
commend that the Y Sensor is removed
from the patient circuit during nebulization.
patient circuit during nebulization.
Nebulization
ebulization
bulization
ulization
lization
ization
zation
ation
tion
ion
on
n