213
Servo
…
User´s manual
US edition
Order No: 66 00 261
Infant
Adult
Universal
Options
9
O
2
Sensor (if installed)
f installed)
No regular overhaul required.
O
2
cell (if installed)
cell (if installed)
Regularly check the cell status in the menu
k the cell status in the menu
Status. The O
2
cell should be exchanged if
be exchanged if
<10% is indicated for it in the
s indicated for it in the Status menu.
Refer to page 177 for O
ge 177 for O
2
cell adaptation.
WARNING! The sealed unit of the O
he sealed unit of the O
2
cell
contains a caustic liquid which may cause
may cause
severe burns to the skin and eyes. In case of
. In case of
contact, immediately flush continuously with
water for at least 15 minutes and seek
15 minutes and seek
medical attention especially if the eyes are
f the eyes are
affected.
Important:
ant:
• Make sure the O
2
cell is for the Servo-i
or the Servo-i
model. O
2
cell package must have a blue
ge must have a blue
label.
• Replacement of the O
of the O
2
cell and filter and
extended cleaning must only be
st only be
performed by personnel who are well
trained in its use.
use.
Bacteria filter for O
2
cell
The bacteria filter for the O
2
cell can be
exchanged when the cell is replaced, or at
ged when the cell is replaced, or at
other intervals according to hospital routine.
SVX-9044
SVX-9050
General
eneral
neral
eral
ral
all