Let the parts soak in a disinfectant agent
gent
such as
– Alcohol (ethyl- or isopropyl alcohol))
– CidexOPA
– Hexanios G+R, Aniosyme DD1 and
G+R, Aniosyme DD1 and
d
Anioxide 1000.
Important: Follow the disinfectant agent
ant agent
manufacturer’s recommendations and
instructions. Otherwise the cassette may be
wise the cassette may be
damaged.
Rinse after disinfectant
disinfectant
• Rinse the parts thoroughly in water to
nse the parts thoroughly in water to
remove all traces of disinfectant. Let the
ces of disinfectant. Let the
water flow through the parts.
parts.
• Rinse the expiratory cassette by dip it in
dip it in
water and carefully shake and tilt the
cassette holding it vertically in both
y in both
directions, repeat this 3-4 times. Mineral
mes. Mineral
deposits on the expiratory cassette affects
cassette affects
the function. It is important to rinse the
expiratory cassette thoroughly. Residues
ly. Residues
from chemicals can affect the patient,
patient,
cause leakage and extra stress on the
material.
Drying
ying
Refer to page 203.