The expiratory cassette
piratory cassette
• When using an external bacterial filter the
filter the
only cleaning needed is wiping the
exterior.
xterior.
• Rinsing the cassette in water (<35°C/95°F)
35°C/95°F)
immediately after use may be a useful
fter use may be a useful
alternative to disinfection. Immediate
rinsing increases the probability of
g increases the probability of
removing particles and minimizes the risk
ving particles and minimizes the risk
of cross contamination between patients.
• After processing, always make sure that
g, always make sure that
the cassette is dry. If not, the cassette may
not, the cassette may
fail the Pre-use check.
• Never dry the cassette by applying high-
pressure air. Internal tubing may be
ure air. Internal tubing may be
damaged.