The Servo Ultra Nebulizer is intended for
zer is intended for
nebulizing drugs to patients requiring
mechanical ventilation or positive pressure
breathing assistance via an endotracheal
dotracheal
tube or face mask / prongs.
prongs.
The nebulizer operates continuously
y
regardless of ventilation mode setting. No
extra gas volume is added to the inspiratory
ume is added to the inspiratory
Minute Volume and thus neither the ventilator
Volume and thus neither the ventilator
settings nor the readings are affected.
cted.
WARNINGS!
RNINGS!
• Servo Humidifier/HME must be
E must be
disconnected during nebulization.
Otherwise the humidifier may be blocked.
ocked.
• The heated humidifier must be switched
witched
off during nebulization. Otherwise the
Otherwise the
particle size may be affected.
ffected.
• The nebulizer must not be used without
without
buffer liquid (sterile water). Otherwise the
Otherwise the
ultrasonic generator crystal may break.
may break.
• When a Servo Ultra Nebulizer is used,
Ultra Nebulizer is used,
always consult the drug manufacturer
onsult the drug manufacturer
regarding the appropriateness of
ultrasonic nebulization for certain
bulization for certain
medications.
• When the ventilator is running on batteries
the nebulizer module is inoperative to
module is inoperative to
reduce the power consumption.
• During nebulization a filter must be
connected to the expiratory inlet of the
d to the expiratory inlet of the
ventilator. Always carefully monitor the
s carefully monitor the
airway pressure during nebulization.
during nebulization.
Increased airway pressure could be
pressure could be
caused by a clogged filter. The filter should
he filter should
be replaced if the expiratory resistance
increases or every 24 hours when the
very 24 hours when the
nebulizer is being used.
g used.
Note:
• For information about the stand alone
Aeroneb Professional Nebulizer System,
ystem,
refer to separate manual.
• The
Y Sensor measurement (page 135)
35)
can be incorrect when the Aeroneb
Professional Nebulizer System is in use.
fessional Nebulizer System is in use.
Therefore, we recommend that the Y
Sensor is removed from the patient circuit
during nebulization.