239
Servo
…
User´s manual
US edition
Order No: 66 00 261
Infant
Adult
Universal
Options
10
0
Error code no.
Possible cause/ part of sub
part of sub
system
Remedy
Technical error no. xxxx
xxxx
(General)
Technical problem, identified by the
error code xxxx.
Restart the ventilator and perform a
ventilator and perform a
Pre-use check. If the problem still
remains, take the unit out of
operation and contact a service
technician.
Technical error no. 1 - 6,
29, 10001
0001
Power failure.
wer failure.
Contact a service technician.
Technical error no. 7, 10-
11
Expiratory / Inspiratory channel
Inspiratory channel
failures.
Contact a service technician.
Technical error no. 12,
16
Connection failures.
Contact a service technician.
Technical error no. 25,
43
3
Communication failure.
Contact a service technician.
Technical error no. 27
Test of back-up sound device failed.
k-up sound device failed. Restart the ventilator and perform a
Pre-use check. If the problem still
blem still
remains, take the unit out of
operation and contact a service
technician.
Technical error no. 28,
20004
Alarm sound level too low.
w.
Check that the loudspeaker outlet is
not obstructed.
Restart the ventilator and perform a
form a
Pre-use check. If the problem still
remains, take the unit out of
operation and contact a service
technician.
Technical error no. 38-39 Barometer failures.
Contact a service technician.
Technical error no. 8-9,
33-35, 41
5, 41
Timeout failures.
Contact a service technician.
Technical error no. 46
Internal failure, Alarm output
put
circuitry.
Contact a service technician.
Technical error no. 48
Timeout failures.
Contact a service technician.
Technical error no. 49
Timeout failures.
Contact a service technician.
Technical error no. 51
Technical problem with Y Sensor
with Y Sensor
module.
Contact a service technician.
Technical error no.
20002
Backlight broken.
Contact a service technician.
Technical error no.
20003
3
Button stuck.
Check User Interface buttons.
Contact a service technician.
Technical error no.
40001
Exp. flow meter failure.
Contact a service technician.
Technical error no. 22,
24,40, 42, 44, 45, 50
10002-10003, 20001
Various failures.
Contact a service technician.
Technical error messages
echnical error messages
chnical error messages
hnical error messages
nical error messages
ical error messages
cal error messages
al error messages
l error messages
error messages
error messages
rror messages
ror messages
or messages
r messages
messages
messages
essages
ssages
sages
ages
ges
es
s