There are 2 alternatives for calibration: Cell
Cell
zero and Verification.
1. Press the fixed pad
he fixed pad Menu.
2. Press the pad Options.
3. Press the pad CO
2
calibration.
bration.
Contents of
ntents of CO
2
calibration window:
4. Cell zero: The cell zero calibration is
bration is
used when the Capnostat sensor has
been shifted.
5. Verification
fication: The verification calibration
includes cell zeroing, verification against
s cell zeroing, verification against
reference cell and adapter zeroing.
Note: The verification calibration is
bration is
recommended. A verification should always
uld always
be performed when the airway adapter is
the airway adapter is
altered, if a faulty Capnostat sensor is
suspected or when prompted. This ensures
cted or when prompted. This ensures
adequate measurements and calculations.