Suction Support
Loops
Scales
Waveform
configuration
Open Lung
Tool
Open Lung
Tool scales
This function allows the user to disconnect the patient from the
ventilator and turning off the alarms (for apnea, minute volume,
frequency alarm, check tubing, EtCO2, low PEEP) during suction
support.
Note. Suction Support is not available when NIV or O2 Breaths is
activated.
The sweep speed, mm/s, can be set to 6, 10 or 20 mm/s (default).
Note. Open Lung Tool is not available in NIV
Possibility to increase the space viewing the waveform curves.
This in turns means that more detailed information can be gained.
These three scales are by default set to automatic scaling.
Note. A reference loop can be saved by
the user.