The Direct Access Knobs allow for
w for
immediate adjustment of vital parameters:
• O2 concentration
• PEEP-level
• Respiratory Rate
piratory Rate
• Tidal/Minute volume/Pressure above
ute volume/Pressure above
PEEP.
Which vital parameters are available
hich vital parameters are available
depends upon the ventilation mode you are
you are
in and how the ventilator is configured.
The four Direct Access Knob parameters are
parameters are
automatically selected depending on the
active mode of ventilation.
ve mode of ventilation.
1. Turn the knob to the desired value
b to the desired value
shown in the corresponding Set
g Set
parameter box above the knob.
2. If the bar is white, the setting is within
what is generally considered safe limits.
dered safe limits.
3. If the bar turns yellow, the setting is too
w, the setting is too
low or too high compared to what are
generally considered safe limits
y considered safe limits
(advisory information).
4. If the bar turns red, the setting is
he bar turns red, the setting is
significantly outside what are generally
what are generally
considered safe limits (advisory warning,
(advisory warning,
accompanied by an audible signal and
by an audible signal and
an advisory message).
When you come to a safety limit, the knob is
b is
inactive during 2 seconds. This is a safety
his is a safety
precaution, intended to make you aware of
ware of
that you have passed a safety limit.
WARNING!
When you turn a Direct Access
Knob, ventilation will change accordingly.
y.
from the next breath without additional
confirmation.
SVX-6020_XX
Direct Access Knobs