231
Servo
…
User´s manual
US edition
Order No: 66 00 261
Infant
Adult
Universal
Options
10
0
CO
2
sensor error
Hardware error in CO
2
Capnostat
stat
sensor. The values in the Capnostat
memory failed the internal test.
Check that the Capnostat sensor is
k that the Capnostat sensor is
properly plugged in. Re insert the
Capnostat sensor if necessary.
Calibrate the Capnostat sensor
brate the Capnostat sensor
(refer to page 151). Change the
Capnostat sensor. If the error
persists, contact a service
technician.
CO
2
sensor temperature
too high
high
Possible hardware error. The
Capnostat sensor temperature is
higher than 50
gher than 50
o
C.
Make sure that the Capnostat
sensor is not exposed to extreme
me
heat (heat lamp, incubator etc.). If
the error persist, the problem is
most likely a faulty Capnostat
Capnostat
sensor.
CO
2
sensor temperature
too low
The Capnostat sensor does not
reach operating temperature.
The problem is probably hardware
y hardware
related. Change Capnostat sensor
and/or module. Call a service
ce
technician.
CPAP High/Low
PAP High/Low
Preset or default alarm exceeded.
Check patient and breathing
breathing
system. Check mas k/ prongs size
and accurate adaptation to the
d accurate adaptation to the
patient
Check alarm settings.
etCO
2
high
Hypoventilation. Leakage with high
ypoventilation. Leakage with high
bias flow. CO
O
2
sensor, Y-piece,
HME
Check patient circuit. Ventilator
tient circuit. Ventilator
settings.
etCO
2
low
Hyperventilation. Leakage with high
h high
bias flow. CO
2
sensor, Y-piece,
HME
Check patient circuit. Ventilator
settings.
gs.
Exp. cassette exchanged Expiratory cassette has been
exchanged during operation. No
Pre-use check performed after
exchange.
Perform a Pre-use check.
Inspiratory flow
low
overrange
Combination of settings exceeds the
allowable inspiration flow range.
Change the ventilator settings.
ventilator settings.
Increase the gas inlet pressure.
Internal temperature:
High
Temperature inside the ventilator is
too high.
Check the function of the fan.
k the function of the fan.
Check the operating temperature.
Nebulizer disconnected
The nebulizer is disconnected
s disconnected
during nebulization. Technical
problem with connection cable.
Connect the nebulizer or change
connection cable.
Nebulizer hardware error
Technical problem with nebulizer
hardware. Temperature too high.
dware. Temperature too high.
Not enough buffer liquid. Technical
quid. Technical
problem with connection cable.
Restart the nebulizer.
start the nebulizer.
Check buffer liquid level. Change
the nebulizer. Change connection
cable. If the problem still remains
If the problem still remains
contact a service technician.
Nebulizer inhibited due
to overheating
g
Temperature too high.
Turn off the nebulizer and restart
when cooled down.
Problem
(Display message)
play message)
Possible cause
Remedy
Medium priority alarms
edium priority alarms
dium priority alarms
ium priority alarms
um priority alarms
m priority alarms
priority alarms
priority alarms
riority alarms
iority alarms
ority alarms
rity alarms
ity alarms
ty alarms
y alarms
alarms
alarms
larms
arms
rms
ms
s