1. The ventilator starts in control mode and
de and
operates according to the Volume
Control, PRVC or Pressure Control
mode. If the patient triggers a breath, the
he patient triggers a breath, the
ventilator will turn to support mode, to
de, to
encourage the patient's respiratory
drive.
2. If the patient is breathing adequately:
y:
a. In Volume Support the ventilator
adjusts the inspiratory pressure level
djusts the inspiratory pressure level
breath-by-breath to assure the preset
target volume.
get volume.
b. In Pressure Support the ventilator
he ventilator
assures that the preset inspiratory
pressure level is maintained
constantly during the entire
g the entire
inspiration.
3. Exceeding the default or manually set
ult or manually set
trigger timeout limit without a sufficient
patient effort will cause:
a. In Volume Support; a PRVC or Volume
Support; a PRVC or Volume
controlled breath will be delivered
according to the selected automode
mode
functionality.
b. In Pressure Support; a Pressure
Support; a Pressure
controlled breath will be delivered.
will be delivered.
4. The ventilator initially adapts with a
pts with a
dynamic trigger timeout limit. This means
ut limit. This means
that for the spontaneously triggering
patient, the trigger timeout limit increases
creases
successively until the set trigger timeout
limit is reached.