Gocator Device Files • 125
Gocator 1100 & 1300 Series
Element
Type
Description
CurrentMax
64f
Setting for maximum output current (mA).
CurrentInvalid
64f
Setting for invalid output current (mA).
CurrentInvalidEnable
32u
0 – Output keeps currently value if measurement is invalid.
1 – Outputs CurrentInvalid if measurement is invalid.
DataScaleMin
64f
Setting for measurement value associated with the minimum current.
DataScaleMax
64f
Setting for measurement value associated with the maximum current.
Value
32u
Setting for selected value source.
Decision
32u
Setting for selected decision source.
CurrentLimitMin
64f
Constraint for minimum output current (mA).
CurrentLimitMax
64f
Constraint for maximum output current (mA).
ValueOptions
String
Constraint for eligible value sources (comma-delimited list).
DecisionOptions
String
Constraint for eligible decision sources (comma-delimited list).
Event
32s
Setting for which event control the output:
1 – Measurement
2 – Software
ScheduleEnable
32u
Setting for scheduled output mode. When unscheduled, output updates
immediately. When scheduled, output updates according to a target
value in software command, or a delay.
0 – Not scheduled
1 – Scheduled
Delay
64f
Setting for output delay. The delay is measured from exposure (first
exposure for multiple exposure) to when output is scheduled. Ignored
when ScheduleEnable is 0. The units depends on SystemDomain.
The delay specifies the time or position at which the analog output activates. Upon activation, there is an
additional delay before the analog output settles at the correct value.
d
igital
o
utPut
A DigitalOutput element defines settings for a digital output. There are two DigitalOutput elements, each
identified by a unique id attribute (0 and 1):
<DigitalOutput id=”0”>
DigitalOutput Child Elements
Element
Type
Description
PassMode
32u
Setting to specify how the state of the output is defined:
0 – Pass if decision is true
1 – Pass if decision is false
2 – Pass always
PulseWidth
32u
Setting for digital pulse width (us).
Decision
String
Setting for selected decision sources (comma-delimited list).
PulseWidthMin
32u
Constraint for minimum pulse width (us).
PulseWidthMax
32u
Constraint for maximum pulse width (us).
DecisionOptions
String
Constraint for eligible decision sources (comma-delimited list).
SignalType
32s
Setting for signal type:
0 – Pulsed output
1 – Continuous output
Summary of Contents for Gocator 1100 Series
Page 1: ...USER MANUAL Gocator 1100 1300 Series Version 3 6 5 15 Revision C...
Page 187: ...Specifications 187 Gocator 1100 1300 Series Envelope 21 0 40 0 20 0 71 0 2 1 19 3 101 4...
Page 189: ...Specifications 189 Gocator 1100 1300 Series Envelope 0 39 0 87 5 120 7 0 1 2 11 4 40 0 60 0...
Page 191: ...Specifications 191 Gocator 1100 1300 Series Envelope 21 0 182 5 35 0 72 2 102 8 2 8 19 6...
Page 193: ...Specifications 193 Gocator 1100 1300 Series Envelope 0 39 0 88 9 122 2 0 11 8 1 6 182 5 217 5...
Page 196: ...Specifications 196 Gocator 1100 1300 Series Envelope 21 0 200 0 200 0 72 3 103 1 2 8 19 7...
Page 199: ...Specifications 199 Gocator 1100 1300 Series Envelope 0 39 0 88 9 122 6 0 12 0 1 6 200 0 400 0...
Page 202: ...Specifications 202 Gocator 1100 1300 Series Envelope 21 0 237 5 325 0 72 3 103 2 2 8 19 8...
Page 205: ...Specifications 205 Gocator 1100 1300 Series Envelope 0 39 0 89 0 122 7 0 12 0 1 7 237 5 562 5...
Page 208: ...Specifications 208 Gocator 1100 1300 Series Envelope 21 0 375 0 562 5 72 6 103 3 3 0 19 9...
Page 211: ...Specifications 211 Gocator 1100 1300 Series Envelope 0 39 0 89 0 122 7 0 12 0 1 7 237 5 562 5...
Page 214: ...Specifications 214 Gocator 1100 1300 Series Envelope 400 0 21 0 250 0 72 4 103 3 2 8 19 8...
Page 217: ...Specifications 217 Gocator 1100 1300 Series Envelope 0 39 0 89 0 122 7 0 12 0 1 7 250 0 650 0...
Page 220: ...Specifications 220 Gocator 1100 1300 Series Envelope 21 0 500 0 2000 0 200 1 231 4 2 7 19 9...