Installation
73
Finalize the Installation
IMPORTANT: Before Sealing the LCR 600
After correctly powering up the LCR 600, continue on to the LectroCount LCR 600 Setup and
Operation manual to setup the LCR 600 for operation. We recommend that you setup and test the
LCR 600 before closing and sealing the unit.
Close and Seal the Unit
After setting up and testing the unit, complete the installation by closing and sealing the housing. The LCR
600 must be environmentally sealed to protect the electronics against the elements. The LCR 600 must
also be sealed by a Weights & Measures representative to guarantee that the register is operating at the
proper regulatory standards.
1. Secure the cables behind the LCR 600 and the cables in the cab with cable ties.
2. If conduit was used during the installation, fill the conduit end inside the LCR 600 with silicon RTV
(provided with shipment, PN 82575). Read and follow the directions of the Environmental Sealing
Guidelines, below.
3. Tighten the cover bolt using a large flat-tip screwdriver or a 7/16” socket. Completely close the gap
between the cover and the housing tab. Read and follow the directions of the Environmental
Sealing Guidelines, below.
4. Seal the cover bolt and the switch plate bolts with a wire/lead seal. See the Weights & Measures
Seals section below.
Environmental Sealing Guidelines
The LCR 600 includes sensitive electronic components, including a micro-controller that can be damaged
by the presence of moisture. Therefore, it is essential that all conduit ports, the cover, and the shaft seals
be adequately sealed by the installer to ensure watertight integrity. The conformal coating on the board
and a moisture-absorbing desiccant inside the enclosure mitigates the problem of corrosion due to
moisture, but these measures only protect the board from small amounts of moisture trapped inside when
the lid is closed in humid conditions. They are not adequate in protecting the unit over time if a
continuous leak is present in the enclosure.
Sealing the LCR 600 is the Responsibility of the Installer
There is no product warranty coverage for any water or moisture damage to the LCR 600 that results
from improper sealing.
Summary of Contents for LectroCount LCR 600
Page 1: ...LectroCount LCR 600 E3708 E3709 Series...
Page 13: ...Introduction 13 Dimensions...
Page 14: ...Introduction...
Page 42: ...Installation Disconnect Power Disconnect the power before working on the CPU board...
Page 54: ...Installation...
Page 61: ...Installation 61...
Page 63: ...Installation 63...
Page 72: ...Installation Epson Printer Power Connections LectroCount LCR 600 Power Connections...
Page 75: ...Installation 75 Bill of Materials External Components...