Installation
41
8.
Coat the entire probe length with the copper grease provided. Insert and recoat the probe 2 or 3
times to provide a uniform coating inside the Thermowell and to ensure proper heat transfer from
the liquid to the probe.
9.
Connect the assembled Thermowell to the fitting in the middle of the strainer cover. The angled
fitting at the top of the cover is for weights & measures purposes. See Step 6 above.
10. Connect the conduit to a ½" NPT port on the back of the LCR 600 using the elbow fittings
provided with the conduit kit (PN 81024).
11. Wire the temperature probe to the LCR 600 internal board on connector J14.