Installation
Auxiliary Outputs
Auxiliary Outputs 4 & 5 (LectroCount Remote Display)
Auxiliary outputs 4 & 5 are typically used for the LectroCount Remote Display, but they can also be used
for other types of displays. Signals from these outputs duplicate the volume data sent to the LCR 600
display.
Refer to the LectroCount remote display manual for complete installation instructions.
Follow these steps to wire an external display to the LCR 600:
1. Attach cable glands and/or conduit connectors to the display and the LCR 600 port(s).
2. Thread the wires through a piece of weatherproof conduit cut-to-length from the display port to a
LCR 600 port
3. Pull the wires through the ports and tighten the connectors.
4. Connect the four display terminals to four terminals on the J18 terminal block of the LCR 600 CPU
board.
·
Remote display terminal 20 to LCR 600 terminal 45
·
Remote display terminal 21 to LCR 600 terminal 39
·
Remote display terminal 22 to LCR 600 terminal 40
·
Remote display terminal 23 to LCR 600 terminal 41