Preparation before Surgery
20
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01
7 Preparation before Surgery
7.1 Transport (F20 only)
WARNING
Risk of injury by:
•
Uncontrolled swinging out of swing arm!
•
Tilting of the stand!
•
Feet in non-sturdy shoes may get trapped beneath the
base!
•
Abrupt braking of the surgical microscope at a threshold
that cannot be crossed!
X
X
For transportation, always move the Leica M525 F20
surgical microscope into the transport position.
X
X
Never move the stand while the unit is extended.
X
X
Never roll over cables lying on the floor.
X
X
Always push the Leica M525 F20 surgical microscope;
never pull it.
CAUTION
Surgical microscope can move without warning!
X
X
Always lock the footbrake when you are not moving the
system.
Returning the Leica M525 F20 to the transport position
X
X
Pull the stop lever (1) and rotate it into a vertical position.
X
X
Grasp the handles.
X
X
Move the swing arm up and down until the transport lock
engages.
X
X
Engage all brakes.
X
X
Bring swing arm into transport position.
X
X
Step on the footbrake release levers (3) to release the
footbrakes (2).
X
X
Move the Leica M525 F20 using the handle.
1
2
3