Annex
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01 57
14 Annex
14.1 Checklist before the operation
Patient
..................................................................................
Surgeon
..................................................................................
Date
..................................................................................
Step Procedure
Details
Checked / Signature
1
Cleaning the optical accessories
X
X
Check the tubes, eyepieces and the documentation
accessories (if used) for cleanliness.
X
X
Remove dust and dirt.
2
Installing the accessories
X
X
Lock the Leica M525 F20/CT20 in place and install all
accessories on the microscope so it is ready for use
(see chapter 7.3).
X
X
Connect the footswitch if used.
X
X
Check the camera image on the monitor and realign if
necessary.
3
Checking the tube settings
X
X
Check the tube and eyepiece setting for the selected user.
4
Balancing
X
X
Balance the Leica M525 F20/CT20 (see chapter 7.4).
X
X
Turn the handles forwards and hold.
All breakes are released.
X
X
Check the balancing.
5
Function check
X
X
Connect the power cable.
X
X
Switch on the microscope.
X
X
Switch on the illumination.
X
X
Check the Main Light 1 and Main Light 2 before surgery.
Make sure that the remaining life time is sufficient for the
planned surgery.
X
X
Replace defective bulbs before the surgery.
X
X
Test the operational readiness of the magnification motor
and the focus motor.
X
X
Test all functions on the handles and the footswitch.
X
X
Check the user settings on the control unit for the selected
user.
6
Positioning at the OP table
X
X
Position the Leica M525 F20/CT20 on the OP table as
required and lock the foot brake (see chapter 7.9).
7
Sterility
X
X
Fit sterile components and sterile drape if used
(see chapter 7.3).
8
Final work
X
X
Check that all equipment is in its proper position
(all covers fitted, doors closed).