Operation
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01 31
8.7 Attaching sterile controls
Sterilizable covers can be fitted on the rotary knobs on the Leica
M525 F20/CT20 surgical microscope.
Covers for rotary knobs
Also attach the covers if you are using sterile disposable
drapes; the controls will be easier to grasp.
X
X
Slide steam-sterilizable covers (1) over the adjusting knob for
the field diameter and the adjusting knob for the working
distance.
X
X
Attach steam-sterilizable covers to accessories as well (if
present).
Cover for footswitch
WARNING
Risk of infection!
X
X
Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
Packaging the footswitch in a plastic bag protects it against
dirt.
Sterile drape for stand
General notes:
• If the entire system is draped with a sterile wrap all the way to
the floor, a controlled shutoff of the illumination must be
expected due to overheating.
• Follow the instructions provided by the manufacturer of the
sterile drape.
2
X
X
Switch on main illuminator.
X
X
Release the brakes and extend the swing arm.
X
X
Put on sterile gloves.
X
X
Attach all the sterile controls.
X
X
Unpack the sterile drape carefully and drape it over the Leica
M525 F20/CT20 surgical microscope as far as the swing arm.
Make sure that the ventilation openings (2) are not covered.
X
X
Secure protective glass at objective.
X
X
Do not attach the sterile drape too tightly with the provided
ribbons. It must still be easy to move the instrument.
X
X
Check the ease of movement of the instrument.
X
X
Carefully fasten the sterile drape by the handles.