7.13 Control parameters
The control parameters are optimized ex-works for operation as a bath thermostat (with water as the
heat transfer liquid) with internal control. The parameters are also preset for the operation of external
containers with external control. Sometimes however, the operation of external containers requires ad-
aptation. In addition, the thermal capacity and viscosity of the heat transfer liquid sometimes require ad-
aptation.
The intelligent menu guidance with the Master and Command detects whether you
have set the device (as described in Section 7.7.4), to internal or external control and
only displays the relevant dialog boxes in each case.
Your Proline Thermostat automatically optimizes some control parameters. This au-
tomatic mechanism should only be deactivated and manually optimized in exceptional
cases.
7.13.1 Internal control variable (integral measurement probe)
Only read further here, if you have no external temperature probe connected (and activated according
to Section 7.7.4 as control variable).
Master
P1d..
3x
to
MNEnu
1x
to
P1d..
As shown in the menu structure (
7.6), the parameters for the in-
ternal control variable can be set.
Select the parameter with
or
and confirm with
. The set
value is displayed. Adapt it with
or
and confirm with
.
Example: Proportional range is 8.0 K.
Proportional range
_P
= Xp
in Kelvin.
Reset time
tn
= Tn
in seconds.
Derivative time (Auto/Man)
tv
= Tv
in seconds.
The thermostat logic system only permits values with Tn > Tv!
Damping time (Auto/Man)
td
= Td
in seconds.
Tv, Td changeover to auto/ man
tvd
A
=Auto or
MN
= Manual.
The works setting is Automatic. Only experienced control techni-
cians should change these two parameters!
_P (0
°C
tud A
°C