Command
Explanation
RMP_IN_03
Read the current programme run
RMP_IN_04
Read which program further commands refer
RMP_IN_05
Read which programme is currently running (0 = none)
LOG_IN_00_XXXX
Read measuring point XXXX from data logger
(Reply: e. g. 020.00_021.23_030.50
set point temperature = 20.00
°C, bath temperature = 21.23 °C, external temperature = 30.5 °C)
LOG_IN_01
Read all measuring points from data logger
As a difference to
the command “LOG_IN_00”, a tabulator is used here
as separator instead of ,_’ . The measuring points are separated by CR
and LF. The end is marked by CR LF CR LF.
LOG_IN_02
Read starting time of the data logger
(Reply: e.g. 20_14_12_20
day 20, 14:12:20
o’clock)
LOG_IN_03
Read acquisition interval from the data logger
(Reply in seconds)
For ”_“ use also ” ” (blank character) is also admissible.
The equipment response is always in the fixed decimal format ”XXX.XX“ or for
negative
values “-XXX.XX” or ” ERR_X“. (RS-485 interface e.g. ”A015_ XXX.XX”
or ”A015_-XXX.XX” or ”A015_ERR_X”).
8.3.7 Error messages
Message
Explanation
ERR_2
Wrong input (e.g. buffer overflow)
ERR_3
Wrong command
ERR_5
Syntax error in value
ERR_6
Illegal value
ERR_8
Module or value not available
ERR_30
Programmer, all segments occupied
ERR_31
Set point not possible, analogue set point input ON
ERR_32
TiH
TiL
ERR_33
external sensor missing
ERR_34
Analogue value not available
ERR_35
Automatic is selected
ERR_36
No set point input possible. Programmer is running or is pausing.
ERR_37
No start from programmer possible, analogue setpoint input is switched on.
8.3.8 Driver software for LABVIEW®
An individual, easy-to-use control and automation software for operating the PROLINE device can be
programmed with the aid of the National Instruments program development tool LABVIEW
®
http://sine.ni.com/apps/we/nioc.vp?cid=1381&lang=US
In order to make program operation possible on the RS-232/485 interface, LAUDA provides drivers
specially designed for LABVIEW
®