7.11 Ramp function
With the ramp function temperature changes over any time period can be conveniently entered. This is
especially advantageous with very low temperature changes (e.g. 0.1 °C per day).
Example: From the current outflow temperature (e.g. 242.4 °C), 200 °C of cooling is to occur over
5 days. Then the temperature change is entered as 200 °C and the time as 5 days.
The ramp function is executed until it is manually terminated or until the temperature
limits T
il
(min) or T
ih
(max) described in Section 7.8.3 are attained.
Command
Ramp function
Open the device parameter menu via the
soft key
Menu .
With the cursor keys continue to:
Programmer
Ramp function
.
Confirm with the key
.
Enter a positive or negative temperature
value with
Temp. change
.
With
Time
enter a figure (without time
unit).
With Time unit choose between
Second(s)
up to
Day(s)
.
Under Status the ramp is started
Start
or stopped
Stop
.
When the ramp function is being execut-
ed,
Ramp active
appears in the window
bar.
Without manual switch-off the ramp ter-
minates at the latest at T
il
(min) or T
ih
(max).
Pump Menu
End
T
set
T
fix
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Program 1
Program 2
Program 3
Program 4
Program 5
Ramp function
Pump Menu
End
T
set
T
fix
Status
Temp. change
Time
Time unit
Second(s)
Minute(s)
Hour(s)
Day(s)