Proline Edition X heating thermostats
92
Interface modules
10/9/2018
8.3.4 Protocol RS-485
The interface operates with 1 stop bit, no parity bit and 8 data bits.
Transfer rate either: 2400, 4800, 9600 (Factory setting) or 19200 baud as se-
lected.
The RS-485 commands are always preceded by the device address. There is
provision for 127 addresses. The address must always have 3 digits.
(A000_...to A127_...).
The command from the computer must be terminated with CR.
The response of the thermostat is always terminated with CR.
CR = Carriage Return (Hex: 0D)
Example:
Transfer of setpoint 30.5 °C to the thermostat with address 15.
Computer
Thermostat
„A015_OUT_SP_00_30.5“CR
„A015_OK“CR
8.3.5 Write commands (Data commands to the thermostat)
Command
Explanation
OUT_PV_05_XXX.XX
External temperature given via interface
OUT_SP_00_XXX.XX
Set value transfer with max. 3 places before the decimal point
and max. 2 places after it
OUT_SP_01_XXX
Pump power level 1 to 8
OUT_SP_02_XXX
Operation mode cooling (0 = OFF / 1 = ON / 2 = AUTOMATIC)
OUT_SP_04_XXX.X
TiH outflow temperature limit, high limit
OUT_SP_05_XXX.X
TiL outflow temperature limit, low limit
OUT_PAR_00_XX.X
Setting of control parameter Xp
OUT_PAR_01_XXX
Setting of control parameter Tn (5
— 180 s; 181 = Off)
OUT_PAR_02_XXX
Setting of control parameter Tv
OUT_PAR_03_XX.X
Setting of control parameter Td
OUT_PAR_04_X.XX
Setting of control parameter KpE
OUT_PAR_05_XXX
Setting of control parameter TnE (5
— 979 s; 980 = Off)
OUT_PAR_06_XXX
Setting of control parameter TvE (0 = Off).
OUT_PAR_07_XX.X
Setting of control parameter TdE
OUT_PAR_09_XXX.X
Setting of the correction limitation
OUT_PAR_10_XX.X
Setting of control parameter XpF
OUT_PAR_14_XXX.X
Setting of the setpoint offset
OUT_PAR_15_XXX
Setting of control parameter PropE
OUT_MODE_00_X
Keyboard Master: 0 = free / 1 = locked (corre
sponds to “KEY”)
OUT_MODE_01_X
Control: 0 = internal / 1 = external Pt100 / 2 = external Analogue
/ 3 = external Serial
OUT_MODE_03_X
Keyboard Command Remote Control: 0 = free / 1 = locked
OUT_MODE_04_X
Setpoint offset source: 0 = normal / 1 = external Pt /
2 = external analog / 3 = external serial