Proline Edition X heating thermostats
94
Interface modules
10/9/2018
Command
Explanation
IN_PAR_06
Read control parameter TvE (0 = OFF)
IN_PAR_07
Read control parameter TdE
IN_PAR_09
Read value of correction limitation
IN_PAR_10
Read control parameter XpF
IN_PAR_14
Read setpoint offset
IN_PAR_15
Read control parameter PropE
IN_DI_01
State of contact input 1: 0 = open/ 1 = closed
IN_DI_02
State of contact input 2: 0 = open/ 1 = closed
IN_DI_03
State of contact input 3: 0 = open/ 1 = closed
IN_DO_01
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed
IN_DO_02
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed
IN_DO_03
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed
IN_MODE_00
Keyboard Master: 0 = free / 1 = locked
IN_MODE_01
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analogue / 3 = ext. Serial
IN_MODE_02
Standby operation: 0 = Device ON / 1 = Device OFF
IN_MODE_03
Keyboard Command Remote Control: 0 = free / 1 = locked
IN_MODE_04
Setpoint offset source: 0 = normal/ 1 = ext. Pt/ 2 = ext. analogue/ 3 =
ext. serial
TYPE
Read device type (response e.g. "P 8")
VERSION_R
Read software type of control system
VERSION_S
Read software type of protection system
VERSION_B
Read software type of Command
VERSION_T
Read software type of cooling system
VERSION_A
Read software type of analogue module
VERSION_V
Read software type of RS-232/485 module
VERSION_D
Read software type of digital module
VERSION_M_0
Read software type of solenoid valve (cooling water)
VERSION_M_1
Read software type of solenoid valve (automatic refilling)
VERSION_M_3
Read software type of solenoid valve (shut-off valve 1)
VERSION_M_4
Read software type of solenoid valve (shut-off valve 2)
VERSION_M_5
Read software type of high temperature cooler
STATUS
Read equipment status 0 = OK, -1 = error
STAT
Read error diagnosis response:
XXXXXXX
X = 0 no error, X = 1 error
1. char = fault
2. char = alarm
3. char = warning
4. char = over temperature
5. char = low bath level
6. char = high bath level (at adjustment alarm)
7. char = no external control variable
RMP_IN_00_XXX
Read a programme segment XXX
(response: e. g. 030.00_010.00_005.00_001.00 > set point tempera-
ture 30.00 °C, time = 10 min, tolerance = 5.00 K, pump level = 1)
RMP_IN_01
Read the current segment number
RMP_IN_02
Read the set number of programme runs