Proline Edition X heating thermostats
YACE0099 / 10/9/2018
81
Step 1
– 2 (see above) must follow.
Set the overtemperature cut-off higher than the bath temperature again
and wait until
tEMNP
appears in the display.
Unlock with the
key.
Unlocking is not possible on the Command remote control!
Command
Overtemperature alarm!
Overtemperature alarm! is shown in the display and signifies that un-
locking is only possible on the Master control panel.
7.14.2 Low-level alarm and low-level checking
If the liquid level falls so far that the heating element is no longer com-
pletely covered with liquid, an alarm is initiated:
1. The alarm sounds as a dual-tone signal.
2. Display for
LEUEL
(low level) is shown when the bath contains
too little liquid.
3. The red LED
above the fault triangle
flashes.
Heater switches off on both poles,
Varioflex pump is switched off.
Find the cause of the fault and, where necessary, top up for missing
heat transfer liquid (
Press the Enter key.
Also press this key if the unit has been switched off in the fault
state.
Checking the safety system at regular intervals
by lowering the bath level. To do
this, push hose onto pump connector and pump heat transfer liquid into a suitable
vessel.
Step 1
– 2 must follow.
With this test the bath temperature must not be below 0 °C or above max. 50 °C,
otherwise there is a risk of burning!
If irregularities arise during the checking of the safety devices, switch off the unit im-
mediately and pull out the mains plug!
Have the equipment checked by LAUDA Service Constant Temperature Equipment!
Command
Low-level alarm!
Low-level alarm is shown in the display and signifies that unlocking is only
possible on the Master control panel.
Level alarm
LEUEL
°C