Proline Edition X heating thermostats
YACE0099 / 10/9/2018
65
Using the cursor keys move the black background to the field which you would like to change.
It can be edited by pressing the key
(see following pages).
The soft key Insert inserts in the marked line a new segment which has a default value
taken from the previous segment with the exception of the tolerance field. The tolerance is al-
ways specified as 0.00. All following segment lines will be moved one line downwards.
In the above window Segment 1 was created in this way.
Continue with
to the fields
"Time
”
"Tolerance". See program example in (
If there is no entry in the "Time" field, the bath temperature is approached as quickly as possi-
ble. With a time entry the final temperature is obtained exactly after the time expires (ramp).
The entry in the field "Tolerance" field defines how accurately the final temperature is to be ob-
tained before the next segment is processed. In case there is no additional cooling, you should
select a more generous tolerance limit. Check and observe the transient effect using the
“graphic display” (
Section 7.9).
If the tolerance range has been selected too small, it may be that the program does not
continue, because the required tolerance is never achieved.
External temperature control: Especially with ramps, a too close tolerance range can
cause undesired delays in the start phase of the ramp.
No.
Pump
Out 1
Out 2
Out 3
Start
-------
------- -------
-------
1
4
------- -------
-------
Pump
Menu
End
Insert Delete
Then continue with
to the pump and
signal output setting.
The right-hand part of the entry table ap-
pears as shown on the left.
Here, in the "Pump" field, the pump level
and, in the fields "Out 1" to "Out 3", the
contact outputs of the contact mode (ac-
cessory) can be programmed. With the
setting "-------" the starting value is re-
tained which was either set before the
program start or was defined by a previ-
ous segment in the running program.
Further details are given on the following
pages.