Proline Edition X heating thermostats
YACE0099 / 10/9/2018
55
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the correc-
tion where applicable!).
Change the display in the adjacent win-
dow to the true value with cursor or soft
keys and accept with
or End ,
or quit the window with
without making
changes.
7.8.8 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
dEF 1
3x
to
MNEnu
2x
to
CAL..
1x
Call offset works setting for the internal temperature probe
dEF 1
,
dEF 1
is displayed.
3 seconds
The works setting is accepted after pressing for 3 seconds and
donE
is displayed.
Command
Default
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to:
Settings
Calibration
intern Pt100
Default
.
The window shown adjacent appears.
Confirm selection with
.
dEF 1
°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temperature value of the ref.
temp. measurement device:
20,
15
Min: -50,0°C Max: 302,00
Pump
Menu
End
T
set
T
f
x
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for PROLINE Edition X P 12 C
Page 2: ......
Page 4: ......
Page 112: ...Proline Edition X heating thermostats 112 YACE0099 10 9 2018 Blank page ...
Page 114: ......
Page 115: ......