Proline Edition X heating thermostats
42
/
10/9/2018
Command
Pump level
Open the device parameter menu via the
soft key Menu .
Change from
Pump
Pump level
us-
ing
.
With
or
you enter the illustrated
window.
Level 5
is active.
Select another pump level with
or
and confirm with
or End ,
or quit the window with without making
any changes.
Standby activation
+
Standby activation
(Pump and heater are deactivated when
the green LED in the lower part of the key
is lit).
Please exercise caution when thermostat is in standby mode.
The following settings/ actions may start the thermostat unintentionally from the standby
mode:
A previously activated timer mode (
7.12), because a started time continues
to run.
"Start" command via interfaces (
7.7.4 Activating external control
An external temperature probe can be connected to the Proline Thermostats. How this is done is ex-
plained in Section 7.7.2. If the set point temperature is to be controlled using this sensor instead of the
internal sensor, the setting can be made here.
Furthermore, control can also occur based on the signal from the analog or serial module (
4.8).
Master
Con
and 2 x
Call the source selection for the control
Con
.
The momentary setting for the source is displayed,
here
1
for internal, because control takes place using the temper-
ature signal from the internal temperature probe.
The source display flashes.
Pump Menu
End
T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
Pump level
Con 1
°C