Proline Edition X heating thermostats
88
Interface modules
10/9/2018
8
Interface modules
8.1
Installing modules
The master and command can be supplemented with further interface modules which are simply in-
serted at the back of the control head into two module slots.
Touch the earthed bath cover of the Proline ther-
mostat to discharge any electrostatic charge.
Remove the module from its packaging.
Switch off the thermostat and pull out the mains
plug.
Insert a screwdriver into the lower recess of the
module cavity and prise up the plastic cover. The
cover can then be pulled off downwards.
Pull out the plug of the bus connecting cable from
the plastic cover.
Plug on the bus connecting cable (red plug onto
red socket).
Insert the module and secure with the two cross-
head screws.
Connect the mains plug again and switch on the
thermostat.
The plugs are protected against reverse polarity.
The plugs have a ridge, which slides into a groove
in the socket.