Rossmax Pencil Thermometer TG380 

Quick Guide 

How to Use 

1. To push the On/O

 bu

on by the display window. The display will read for 2 seconds and beep sound will

be heard.

2. Releasing the on/o

 bu

on, last measured temperature will be shown for about 2 seconds. Then shows

the self-test temperature 37.0°C (98.6°F), and then Lo°C (Lo°F) displays.

3. Wipe or dip the probe with alcohol for disinfec

on. Posi

on the probe in oral, rectal or axillary site.

4. Once the degree sign °C (°F) on the display has stopped 

ashing (usually within 10 to 40 seconds), a Beep

signal  will  sound  for  approx.  10 

mes,  the  measured  temperature  is  shown  on  LCD  window.  The

temperature reading will not change a

er the unit is removed from tes

ng posi

on.

If the temperature <37.8°C (100.0°F), the buzzer will alarm “Bi---Bi---Bi---Bi---”.
If the temperature 

 37.8°C (100.0°F), the buzzer whose frequency is higher will alarm “Bi-Bi-Bi---

-Bi-Bi-Bi--” about 10 

mes rather than ”Bi---Bi---Bi---Bi---”

5. The unit will automa

cally turn o

 in 10 minutes for energy-saving. However, to prolong the life of the

ba

ery, it is suggested to press the On/O

 bu

on to turn the unit o

 once the measurement is  done.

6. Please dispose of any used ba

ery and the thermometer according to the local regula

ons.

Note

: A

er step 2, if the room temperature is greater than 32.0°C (89.6°F ), it will be displayed the room 

temperature. 

Note

: Usage of the probe cover may result in a 0.1°C (0.2°F) di

erence from actual temperature. 

Note

: To clean the probe before and a

er using the thermometer to ensure an accurate reading and avoid 

cross contamina

on approximately. 

Oral Use 

1. Place the probe well under the pa

ent’s tongue.

2. Con

rm the pa

ent to keep his mouth closing for about 20 seconds.

Do not drink hot or cold 

uids, exercise, and smoke or perform other ac

vi

es prior to a reading. These

ac

vi

es will raise or lower temperature readings when compared to your normal, average temperature.

Axillary Use 

1. Tightly place probe in the pa

ent’s armpit for about 30 seconds.

* Please ensure the close contact between the probe and the armpit of the baby or children.