7.10 Programmer (PGM)
Almost any temperature/time profile can be created with the programmer. A desired bath temperature
can be approached as quickly as possible or via a defined ramp. Furthermore, the pump level and the
behavior of the switching outputs can be defined. Five temperature/time programs are provided for free
programming. Each program consists of a number of temperature/time segments. Also included are
details of how often the program is to be executed (loops). The sum of all segments of all programs
may be up to a maximum of 150.
Typical segments are:
Ramp:
If a time is specified, then the segment is a ramp which is described by the target temperature,
i.e. the temperature at the end of the segment, and the duration from the start to the end of the seg-
ment.
Step:
Without any specified time the final temperature is approached as quickly as possible.
Temperature hold phase:
No temperature change (i.e. the temperatures at the start and end of a
segment are the same).
The programmer can be controlled or changed via the RS-232 interface, the timer or switch-
ing contacts.
7.10.1 Program example
Cooling time depends on
type of device, load, etc.
Real program example with 6 segments
No.
T end °C
Time
Tolerance
Nr.
Pump
Out 1
Out 2
Out 3
Start
30,00°C
-------
0,00°C Start
-------
-------
-------
-------
1
30,00°C
00:20
0,10°C
1
2
-------
-------
-------
2
50,00°C
00:20
0,00°C
2
3
-------
-------
-------
3
70,00°C
00:40
0,00°C
3
4
-------
-------
-------
4
70,00°C
00:10
0,10°C
4
2
-------
-------
-------
5
60,00°C
00:30
0,00°C
5
2
-------
-------
-------
6
30,00°C
00:00
0,00°C
6
2
-------
-------
-------
Pump
Menu
End
Insert
Delete Pump
Menu
End
Insert
Delete
°C
80
70
60
50
40
30
20
10
0
20
40
60
80
100
120
t/min
Start
This is Segment No. 2