Check all devices (modules) on availability. You can use the graphic
representation in the hardware test (see § 4.4).
Test the infrared receiver and the remote control with the hardware test
Pump. The moving parts of the rotor should be checked from time to time
for freedom of movement. Open the cover and clean the pump rollers one
by one with a moisturized cloth. Check also if there is nothing around the
spindle of the roller that causes the friction (like a hair or something). Test
the pump with the hardware test (see § 4.4.2).
Test the pressure interface with the hardware test (§ 4.4.3).
Test the EMG interface with the diagnostic program (§ 4.4.4).
Perform a functional test on all investigations, available in the measurement
program. Check if the investigations starts, if the graphs and results are
printed and if the investigation data is saved.