4.4.4
EMG test
Choose the TAB
Pressures / EMG
to test the EMG of the Solar Blue module.
Figure 4.22
Test the EMG signal
The procedure is as follows:
Stick EMG electrodes on your arm as
shown in figure.
Strain your arm muscles and check if the
EMG value increases. The EMG signal
should be approximately 200.
You can change the EMG scale by using
the Up/Down key. The values are: 1, 2,
5, 10, 20, 50, 100, 200, 500, 1000.
To exit the test, click the
Close
button.
Figure 4.23
Test the EMG signal
You can also test EMG with an EMG generator. Connect the EMG generator,
choose an output voltage and check if the EMG signal corresponds.