Chapter 4
Remove air from tube.
Connect the flowmeter via Bluetooth (when the hardware test Bluetooth is
used) or,
Connect the flowmeter via a wired connection to an MMS-bus receptacle
(when the hardware test Solar (USB) is used). Make sure the power
adapter is connected to the Solar Blue module in this case.
Place the catheter in the water container which is positioned on the
flowsensor.
Click the
Calibrate pump
button. You will see the progress bar “Busy
calibrating”. The pump stops automatically after 5 minutes.
When there is no weight increase during calibration, you will see an error
message
(‘No weight increase detected on flowmeter’). Click the
OK
button
to abort calibration.
4.4.3
Pressure test
Choose the TAB
Pressures / EMG
to test the functioning of the pressure
connectors on the Solar Blue module and the connected transducers or
catheters.
Figure 4.21
Test the pressures
Summary of Contents for Solar Blue
Page 1: ......
Page 9: ...8 2008 2020 by MMS B V...
Page 19: ...18 2008 2020 by MMS B V Chapter 1...
Page 47: ...46 2008 2020 by MMS B V Chapter 3...
Page 75: ...74 2008 2020 by MMS B V Chapter 4...
Page 81: ...80 2008 2020 by MMS B V Chapter 5...
Page 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Page 139: ...138 2008 2020 by MMS B V...
Page 157: ...156 2008 2020 by MMS B V Appendix A...
Page 165: ...164 2008 2020 by MMS B V Appendix C...
Page 173: ...172 2008 2020 by MMS B V Appendix D...
Page 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Page 175: ...174 2008 2020 by MMS B V Appendix E...
Page 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Page 198: ......