Chapter 9
General
Applied part class: BF.
Degree of ingress protection: IPX1.
Dimensions: 195 x 195 x 155 mm (l x w x h), no clamp.
Weight:
o
approximately 2,5 kg
Mounting possibilities:
o
Rail 25x10
o
Pole 20 - 40 mm
o
Table top
Enclosure material: ABS, UL94 V2 grade.
Classification/Approval
IEC 60601-1:2005 + CORR. 1:2006 + CORR.2:2007 + A1:2012
Protection Class I
IEC 60601-1-2:2014
MDD, Medical Device Directive 93/42/EEC, Class IIa medical device
ANSI/AAMI ES60601-1:2005 + C1:2009 + A1:2012 + A2:2012
CAN/CSA-C22.2 No. 60601-1:2014
SGS contract no. 710253
Site Requirements
Operating and storage temperature b15 °C and +35 °C (+60 °F to
+95 °F).
Humidity between 30% and 75%, non-condensing.
Pump
Pump head type: 4 roller Watson Marlow.
Control loop: None, open loop / calibration possibility.
Maximum yield: 250 ml/min (open tube).
Minimum yield: 5 ml/min.
Accuracy: ± 5% (on mean value).
Rotation directory: counter-clockwise.
Filling tubes: part number OBS 275/C or 275/D.
Summary of Contents for Solar Blue
Page 1: ......
Page 9: ...8 2008 2020 by MMS B V...
Page 19: ...18 2008 2020 by MMS B V Chapter 1...
Page 47: ...46 2008 2020 by MMS B V Chapter 3...
Page 75: ...74 2008 2020 by MMS B V Chapter 4...
Page 81: ...80 2008 2020 by MMS B V Chapter 5...
Page 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Page 139: ...138 2008 2020 by MMS B V...
Page 157: ...156 2008 2020 by MMS B V Appendix A...
Page 165: ...164 2008 2020 by MMS B V Appendix C...
Page 173: ...172 2008 2020 by MMS B V Appendix D...
Page 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Page 175: ...174 2008 2020 by MMS B V Appendix E...
Page 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Page 198: ......