60
© 2008
– 2020 by MMS B.V.
Chapter 4
4.4.2
Pump Test
Choose the TAB
Pump
to test the pump of the Solar Blue module.
Figure 4.20
Pump test
The procedure is as follows:
Hang a bag (or bottle) of saline on the hook.
Install a pump tube in the pump.
Connect the 'input' side of the tube to the bag or bottle.
Connect the flowmeter via Bluetooth (when the hardware test Bluetooth is
used) or
Connect the flowmeter via a wired connection to an MMS-bus receptacle
(when the hardware test Solar (USB) is used). Make sure the power
adapter is connected to the Solar Blue module in this case.
Place the other side of the pump tube in the water container which is
positioned on the flowsensor. With this configuration, the 'inflow' can be
checked as the water is pumped from the bag. The flow measured by the
flowsensor can be compared to the selected pump speed.
Depending on the problem, select the desired pump speed.
Start the pump by clicking the
Start pump
button.
Summary of Contents for Solar Blue
Page 1: ......
Page 9: ...8 2008 2020 by MMS B V...
Page 19: ...18 2008 2020 by MMS B V Chapter 1...
Page 47: ...46 2008 2020 by MMS B V Chapter 3...
Page 75: ...74 2008 2020 by MMS B V Chapter 4...
Page 81: ...80 2008 2020 by MMS B V Chapter 5...
Page 138: ...2008 2020 by MMS B V 137 APPENDICI WARRANTY AND INDEX...
Page 139: ...138 2008 2020 by MMS B V...
Page 157: ...156 2008 2020 by MMS B V Appendix A...
Page 165: ...164 2008 2020 by MMS B V Appendix C...
Page 173: ...172 2008 2020 by MMS B V Appendix D...
Page 174: ...2008 2020 by MMS B V 173 Appendix E Solar Blue Safety Barriers...
Page 175: ...174 2008 2020 by MMS B V Appendix E...
Page 189: ...2008 2020 by MMS B V Warranty for Solar Blue System...
Page 198: ......