MEASure Subsystem
68
Keysight U2741A Programmer’s Reference Guide
MEASure:FRESistance?
Syntax
MEASure:FRESistance? [<range|MIN|MAX|AUTO>
[,<resolution|MIN|MAX|DEF>]]
This command presets, configures the U2741A for four-wire resistance
measurements with the specified range and resolution, and make a measurement.
The reading is sent to the output buffer.
Parameters
Remarks
– You can allow the U2741A to automatically select the measurement range
using autorange or you can select a fixed range using manual range.
Autorange is convenient because the U2741A decides which range to use for
each measurement based on the input signal. For faster measurements, use
manual range on each measurement as additional time is required for
autorange since the U2741A has to make a range selection.
– Autorange thresholds
– If the input signal is greater than can be measured on the selected range, the
U2741A gives an overload indication of
+9.9E+37
from the remote interface.
– If the
TRIG:SOUR
is set to
BUS
, the
MEAS?
command overwrites the
BUS
trigger
and returns a measurement.
Item
Type
Range of values
Default value
range
Numeric
100
Ω
, 1 k
Ω
, 10 k
Ω
, 100 k
Ω
, 1 M
Ω
, 10 M
Ω
, 100 M
Ω
100
Ω
resolution
Numeric
3.00e–5, 2.00e–5, 1.00e–5, 5.00e–6, 2.00e–6, 1.50e–6
1.50e–6
Lower range at:
<10% of range
Upper range at:
>120% of range
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Page 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...