STATus Subsystem
7
Keysight U2741A Programmer’s Reference Guide
121
STATus:QUEStionable:CONDition?
Syntax
STATus:QUEStionable:CONDition?
This command queries the condition register for the Questionable Data Register
group. This is a read-only register and the bits are not cleared when you read the
register.
Description
This query returns the value of the Questionable Data condition register. The
condition register is a read-only register, which holds the live (unlatched)
questionable status of the U2741A. Reading the Questionable condition register
does not clear it. This query returns the binary-weighted sum of all bits set in the
register.
Remarks
For more information on the Questionable condition register, refer to
Example
The following query reads the Questionable condition register.
STAT:QUES:COND?
Typical Response:
+512
See Also